Air Detector Alarm
Appears:
Air has entered the venous drip chamber.
Machine actions:
-
The blood pump stops.
-
The venous and the arterial blood line clamps closes.
-
The ultra
fi
ltration rate is set to zero and the dialysis
fl
uid is bypassed from the dialyzer (when blood is
detected).
Possible cause:
-
The blood line connections are not properly connected
(at the arterial needle, at the dialyzer).
-
The blood lines or the dialyzer has not been properly
de-aired during priming.
-
Air have entered the blood lines e.g. when an infusion
is connected to the blood lines.
-
The arterial needle is dislocated.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms