Performance and specification -
Supervisory system
Blood Pressure Supervision
Venous pressure
Operating range
-700 to +750 mmHg
Alarm limits, in treatment
mode
10 to 500 mmHg
Alarm limits, in priming
mode
-100 to 500 mmHg
Accuracy
±10 % within range -700 to -500 mmHg
±5 mmHg or ±3 %, whichever is largest
within range -500 to 500 mmHg
±10 % within range 500 to 750 mmHg
Arterial pressure
Operating range
-700 to +750 mmHg
Alarm limits
-700 to +750 mmHg
Accuracy
±10 % within range -700 to -500 mmHg
±5 mmHg or ±3 %, whichever is largest
within range -500 to 500 mmHg
±10 % within range 500 to 750 mmHg
Extracorporeal blood loss to the environment
Detection method
Venous pressure supervision
Air Detection
Detection method
Ultrasonic detector placed at the venous
drip chamber. The detector has a two
channel structure and the function of the
detector is tested at the function test made
by the microcomputers.
Drip chamber size
Diameter 22 mm
Sensitivity
Bubbles larger than 1
μ
l will be trapped
by the drip chamber. An alarm will be
issued if the blood level falls below the
middle of the air detector.
Extracorporeal blood loss due to coagulation
Detection method
Supervision of the stop time of the blood
pump
Alarm
The attention "
BLOOD PUMP
STOP TIME EXPIRED
,
Start blood pump
"
9:14
AK 96
®
Operator's Manual - Technical Data
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...