Note
•
Always follow the manufacturer's recommendations for the
minimum rinsing volume (for the blood lines and the dialyzer).
•
After the machine has passed the function check and the
bypass path on the Flow Diagram lights up green, it is
possible to activate concentrate stand-by mode in order to save
concentrates before patient connection. See information in
"Concentrate Stand-by Mode" on page 3:30 in part 1.
Note
Extra Priming Procedure
If necessary, continue priming by pressing the
Select key
when
EXTRA PRIMING
has been selected. The menu will change and
request the operator to press the
fl
ashing
Blood pump button
when
ready for extra priming. When the button is pressed, the pump
speed will automatically start at 100
ml/min. Adjust as
necessary using the
Blood Pump Up/Down keys
.
The extra priming overview menu, which gives current
information on the most important extra priming parameters,
will be shown on the Information Display. The preset extra
priming volume is shown below
PRESET VOLUME
. Below
ACC VOLUME
, to the left of the slash, the current total priming
volume obtained since priming start is shown. The total preset
priming volume obtained since priming start is displayed to the
right of the slash.
When the extra priming volume
has been achieved, the
blood pump automatically stops and an attention alarm appears.
EXTRA PRIMING
can be restarted as desired. Check priming
fl
uid and waste bag.
Select
CONNECT PATIENT
using the
Keypad
when the extra
priming procedure is complete, and the menu will change.
Continue by following the instructions in "Treatment", page
If requested, it is also possible to select
RECIRCULATION
at
this moment, see the following instructions.
If the operator wishes to restart priming and at the same time set
the priming volume value to zero, select
NEW PRIMING
when
the priming volume achieved menu is displayed, using the
Keypad
. The machine will reactivate priming and it is possible to
repeat the priming procedure from the start.
Recirculation Priming Procedure
If requested, extend priming by a recirculation priming procedure
where the blood circuit is rinsed using the same priming
fl
uid
during the complete procedure.
Before starting the procedure, the extracorporeal blood circuit
must be prepared for recirculation priming in accordance with
4:50
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...