2.c.
Bring down the upper latch and place the top port of
the cartridge into the hole of the upper latch. Press
down the upper latch to puncture the cartridge.
The BiCart cartridge will be primed automatically.
Note
•
Follow the instructions for use of the BiCart cartridge.
•
The BiCart cartridge can be attached before, during or after
the function check. This means that it can be attached after
fi
nalized function check but must be attached to obtain green
bypass path of the Flow Diagram.
Note
3.
Con
fi
rm concentrates
The con
fi
rm concentrate menus will automatically appear
when the red concentrate connector is removed from the
stand-by port.
3.a.
Select
BICARBONATE
using the
Keypad
.
3.b.
Con
fi
rm that the connected concentrates correspond
with the displayed concentrate alternative on the
Information Display, using the
Select key
.
Note that at any time after con
fi
rming the concentrates, it
is possible to enter the concentrate menus to check/select
concentrates. Press the
Fluid path button
and select
CONC
. Go
into the sub menus and check/select the connected concentrates
using the
Keypad
.
4:10
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...