Heparin
From the Blood Path main menu; select
HEPARIN
by using the
Keypad
to reach the sub menu.
The following will be shown on the Information Display:
Type of syringe (e.g. Terumo 30 mL).
BOLUS
; set heparin solution bolus volume.
FLOW RATE
; set heparin solution
fl
ow rate
STOP LIMIT
; set stop time in minutes at which the heparin pump
should stop before treatment end
ACC HEP
; sum of accumulated bolus volume during priming and total
accumulated volume since treatment start
The heparin pump will run concurrent to the blood pump in double
needle treatment. In single needle treatment it will be running
continuously irrespective of if the blood pump is running or not.
Single needle
From the Blood Path main menu; select
SINGLE NEEDLE
by using
the
Keypad
to reach the sub menu.
The following will be shown on the Information Display:
MIN STROKE
; displays the low alarm limit for the stroke volume
PV LOW
; low venous pressure alarm limit
P VEN
; current venous pressure
PV HIGH
; high venous pressure alarm limit
SET/MEAN QB
; displays set and mean blood
fl
ow rates. The
set blood
fl
ow rate (arterial) is to be set/adjusted using the
Blood Pump Up/Down keys
. The mean blood
fl
ow rate is the effective
blood
fl
ow rate displayed when performing single needle treatment
3:14
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...