Continuous measuring of Kt/V and clearance
Procedure
1. Press the
Fluid Path button
,
select
DIASCAN
.
2. Select
SET DIASCAN
and the
setting menu will be displayed.
3. When
DIASCAN OFF
has been
selected, press the
Select key
.
4. Select
15 MIN
,
30 MIN
,
60 MIN
using the
Display Up key
. Continuous
measurement of clearance is now
activated.
5. If an alarm limit for clearance is
required, select
K LIMIT
. Press
the
Select key
and set a suitable
alarm limit value
using the
Display Up key
. For further
information; see "Clearance
alarm limit", page 12:18.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan
12:27