Immunity test
IEC 60601
test level
Compliance level
Electromagnetic enviroment-guidance
Electrostatic
discharge (ESD),
IEC 61000-4-2
±6 kV contact
±8 kV Air
±6 kV contact
±8 kV Air
Floors should be wood, concrete or
ceramic tile. If
fl
oors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst,
IEC 61000-4-4
±2 kV for power
lines
±2 kV for power
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV for
differential mode
±2kV for common
mode
±1kV for
differential mode
±2kV for common
mode
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
AK 96 dialysis machine requires
continued operation during power mains
interruptions, it is recommended that
the AK 96 dialysis machine be powered
from an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic
fi
eld
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fi
elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note
•
U
T
is the a.c. mains voltage prior to application of the test level.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:19
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...