TMP
When the
TRANSMEMBRANE PRESSURE CONFIRM MENU
is
displayed on the Information Display, press the
Select key
to con
fi
rm
the settings i.e. that the current TMP value and the automatically set
TMP alarm limits are set in accordance with the current UF rate and the
UF coef
fi
cient of the dialyzer being used. If the TMP alarm limits have
not been con
fi
rmed within two minutes, an attention appears where the
operator is requested to con
fi
rm the current settings.
If necessary the TMP alarm limits can be manually set/adjusted within
the preset alarm window by
fi
rst pressing the
Fluid Path button
,
selecting TMP, and adjusting the alarm limits using the
Keypad
. For
instructions on manual setting of the alarm limits, see "Fluid Path
button" in chapter 3. Go to "Transmembrane Pressure" last in the list.
The TMP alarm window will automatically be widened when the
UF Start/Stop button
is pressed.
CAUTION
When negative TMP alarm limits are set, the operator will not be
notified via alarm or attention alarm
that backfiltration may occur.
However, it is possible for the authorized technician to preset the
machine so that when negative TMP alarm limits have been set,
such an attention alarm occurs.
CAUTION
Discontinuing
Follow the instructions for discontinuing in "Discontinuing", page 4:59
in part 1, without deactivating the single needle function.
5:12
AK 96
®
Operator's Manual - Single Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...