Dialysis fluid preparation
Temperature
Temperature alarm (
fi
xed)
40 °C (±0.5 °C)
Overheat protection
The heater can only be on if there is a
fl
ow through the heater.
The
fl
ow is detected by a
fl
ow switch.
Conductivity
Alarm limits
±5 % of the calculated conductivity set
value
Accuracy
0.2 mS/cm
Ultrafiltration Supervision
TMP
TMP is de
fi
ned as the difference, P
b out
-
P
d out
, where P
b out
is the venous pressure
and P
d out
is the pressure measured in the
dialysis
fl
uid, where it enters the machine
after the dialyzer.
Alarm limits
-100 to 500 mmHg
Accuracy
±10 mmHg or ±5 %, whichever is largest
(within range ±500 mmHg)
Blood leakage detection
Detection method
Infrared light detector.
Sensitivity
Alarm will be given for
≥
0.3 ml
blood, hematocrit 32 %, per minute at
300-700 ml/min dialysis
fl
uid
fl
ow. Time
delay for alarm, maximum 5 seconds
(diffusion mode).
pH supervision
pH supervision is only available if the optional pH is installed.
pH measurement of
Dialysis Fluid
1 to 9.9 pH units
±0.2 pH units for pH 5 to 9
±0.5 pH units for pH < 5 or pH > 9
Alarm limits
5.0 to 9.0 pH
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:15
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...