Blood Pump
The blood pump parts are:
a
) the Pump Housing
b
) the Pump Shaft
c
) the Pump Cover
d
) the Pump Rotor and
e
) the Pump Handle
as shown in the corresponding
fi
gure.
The blood
fl
ow measuring of the AK 96 dialysis machine
is based on blood pump rotations (see also in "Blood Flow
Control" on page 9:2 in part 1).
If the blood pump cover is opened whilst the blood pump is
running, it will stop until the cover is closed again.
During a power failure the pump can be manually operated
by turning the pump rotor in an anticlockwise direction,
using the pump handle.
The pump cover may be opened
with ease by pressing the
middle of the cover at the same time as the cover is being
opened, as shown in the corresponding
fi
gure.
2:8
AK 96
®
Operator's Manual - Machine Description
HCEN12239 Revision 12.2010
Program version 3.xx