AIR LEAKAGE
Check fluid tube connections
Appears:
When the dialysis
fl
uid tubes are not properly connected
to the dialyzer.
Machine actions:
Buzzer is activated 2 min. after the attention alarm is
generated.
Measures:
Check the dialysis
fl
uid tubes connections. See
instructions in "Dialysis Fluid Tubes - Attach" on page
Mute time:
2 min
Buzzer sound:
Soft buzzer
CAUTION
Incorrect weight loss may occur if the machine is left
for a longer period with this attention alarm.
CAUTION
14:6
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order