UF VOLUME ACHIEVED TOO EARLY
Press SELECT key to confirm
Appears:
When the set UF volume (
SET UF
) has been achieved
and the remaining treatment time is more than 20 minutes.
Machine actions:
The machine has automatically set the current UF rate
to minimum UF rate.
Measures:
Press the
Fluid Path button
, select
UF
to reach the
UF MENU
. Check that the set minimum UF rate,
SET MIN UFR
, is a proper value for the paricular
patient. If not, adjust the value using the
Keypad
. If
necessary, consider adjusting the set treatment time or
UF volume within the limits for the patient´s needs and
tolerance.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:123
Attention alarms in alphabetical order