Measurement result values, Kt/V measurement
included
-
LAST DIASCAN
shows
current time
for when the
Diascan measurement check was carried out.
- The value below
K
shows the last
measured clearance
value in ml/min.
- The value below
Kt/V
shows
accumulated Kt/V
achieved since treatment start.
- A
Kt/V forecast value
is displayed next to the
accumulated Kt/V value. This value shows the Kt/V
value estimated to be obtained when the treatment is
complete. For further information, see "Forecast" in this
section.
- TTT,
time to target
, shows an estimation of the
treatment time needed to reach the desired Kt/V target
value set by the operator. For further information, see
"Time To Target" in this section.
-
LOG
shows the 20 last measurement results. Scroll up
and down in the list using the
Display Up/Down keys
.
Forecast
Forecast is a calculation of the Kt or Kt/V value estimated
to be obtained when the treatment is complete. If
distribution volume (
VOLUME
) has been set, this Kt/V
forecast value will be calculated by the machine and
will automatically be shown as soon as the
fi
rst Diascan
measurement check has been performed. If distribution
volume has not been set, the machine will only calculate
and display the Kt forecast value. The calculated forecast
12:10
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...