Remove tubes from safety bypass
Press SELECT to abort the test
Appears:
It was not possible for the machine to restart the dialysis
fl
uid tube sensor test.
Machine actions:
None.
Measures:
To be able to continue the activated disinfection or rinse
program; manually check that the dialysis
fl
uid tubes are
properly attached to the safety couplings of the machine.
Con
fi
rm this by pressing the
Select key
and the dialysis
fl
uid tube sensor test will be bypassed.
Alternately, consult the authorized technician responsible
for the machine.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:108
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order