6. Select
SINGLE NEEDLE
using the
Keypad
.
7. Activate the single needle function by pressing the
Select key
for 3 seconds. If desired adjust the alarm limit for
minimum stroke volume
MIN STROKE
.
8. Connect the blood lines to the patient as standard procedure.
Remove any obstructing clamps.
9. Start the blood pump.
The connect patient overview menu, which gives current
information on the most important parameters during patient
connection, will be shown on the Information Display. To
monitor the patient's blood access before blood has been detected
in the venous blood line; read off current arterial pressure below
ART PRESS
and current venous pressure below
VEN PRESS
.
The blood
fl
ow rate is displayed as
PUMP SPEED
until blood is
detected in the venous blood line.
Note
•
The blood pump can be automatically stopped
when the
priming detector detects blood in the venous blood line and the
blood path in the Flow Diagram lights up.
•
During patient connection, when blood has been detected in
the venous blood line, the venous clamp can be open
when
the blood pump is stopped (manually, or automatically ).
This preset is to facilitate removal of air from the venous blood
line after connecting the venous blood line to the patient.
Bear in mind that when this preset is active, the operator is
fully responsible for manually clamping the venous blood line
and checking it from air whenever necessary during patient
connection.
Note
10. Set the required blood
fl
ow rate suitable for the patient using
the
Blood Pump Up/Down keys
; the blood
fl
ow menu will be
displayed, adjust
SET QB
. Especially note that the blood
fl
ow rate in
SET QB
and
ACTUAL QB
shown in the blood
fl
ow menu on the Information Display, is the blood
fl
ow rate
when the blood pump is running in the arterial phase of the
single needle cycle. It is
not
the mean blood
fl
ow rate for
the single needle treatment.
5:6
AK 96
®
Operator's Manual - Single Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...