Profiling Models
The pro
fi
ling function of the AK 96 dialysis machine consists of three
different models (memories). Pro
fi
les for sodium (
NA
), bicarbonate
(
HCO3
) and ultra
fi
ltration rate (
UF
) can be set for each model. It is
possible to set pro
fi
les for all three parameters at the same time or
just pick one or two parameters within the model. The models can be
manually set or they can be preset by an authorized technician.
Profiling Setting/Activation
Pro
fi
ling can be started at treatment start or at any time during
treatment. The model that is to be used can be manually set or be preset
by an authorised technician.
Step-by step instructions for the two alternatives are described further
on in this section, in "Pro
ling without Preset Model", page 7:7 or in
ling with Preset Model", page 7:10 in part 1.
Pro
fi
ling settings
can be done as soon as the machine has been
switched on and “FCh” is shown on the Time Display. It is possible
to manually set a model or just select a preset model. Pro
fi
les for
sodium (
NA
), bicarbonate (
HCO3
) and ultra
fi
ltration rate (
UF
) can be
set/adjusted for each model. If UF pro
fi
ling (
UF
) is to be used, the
UF volume for the following treatment has to be set before setting
the UF pro
fi
ling.
The pro
fi
ling function for all three pro
fi
ling parameters can be reached
by
fi
rst pressing the
Fluid Path button
, and then selecting
UF
or
COND
using the
Keypad
.
Activation
The pro
fi
ling function can be activated at the same time as
when the setting is being done, or at any other time desired during
treatment. When UF pro
fi
ling has been activated it is possible to
change the pro
fi
ling settings without
fi
rst deactivating pro
fi
ling.
However, changing sodium and/or bicarbonate pro
fi
ling settings cannot
be done without
fi
rst deactivating pro
fi
ling. The pro
fi
ling function
starts to be active when blood has been detected in the venous line and
the
UF Start/Stop button
has been pressed.
7:6
AK 96
®
Operator's Manual - Profiling
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...