It is possible to override
(bypass) the machine alarm
actions for low venous pressure, low arterial pressure, air
detector and blood leak alarms for a limited amount of
time . The purpose of the override option is to facilitate
the handling of the machine in alarm situations. It is
possible to override the alarm actions for more than one
alarm at the same time, see instructions for each alarm in
the Alarm List in this chapter. If the cause of the alarm
has not been corrected within the override time, the alarm
actions will automatically recur. The override time
can be preset by authorized technician between 5-120
seconds except for the override of the air detector which
can be preset 0-15 seconds.
Alarm History
When the
Alarm button
is pressed for 3 seconds, in a
non-alarm situation, a list of the most recently generated
alarms will be shown on the Information Display. The
last generated alarm is shown at the top of the list. The
arrows placed on both sides of the alarm text shows
where you are in the list, scroll up and down by using the
Display Up/Down keys
of the
Keypad
.
The time for when the alarm was generated is shown to
the left on the Information Display.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:7
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...