Heparin Pump
Automatically heparin administration is only available if the optional
heparin pump is installed.
Heparin pump
fl
ow rate
0 - 10 ml/h (±1 ml/5h or ±5 %) The
accuracy is based on tests with 20 ml and
30 ml syringes with diameter of 20 mm.
Heparin bolus
fl
ow rate
60 ml/h with 20 ml (diameter 20 mm)
and 30 ml (diameter 22 mm) syringes.
Bolus volume
0 to 10ml (±0.2 ml or ±5% whichever is
largest, with blood lines primed)
Size
Syringes shall comply with ISO 7886-2.
Stop time
The heparin pump stops before end of
treatment 0.00 to 9.59 h
Counter pressure
Maximum 400 mmHg
Accumulated volume
0 - 999.9 ml
Blood pressure
Venous Pressure
Operating range
-700 to 750 mmHg
Alarm limits
10 to 500 mmHg in treatment mode
-100 to 500 mmHg in priming mode
Accuracy
±10 % within range -700 to -500 mmHg.
±5 mmHg or ±3 %, whichever is largest
within range -500 to 500 mmHg.
±10 % within range 500 to 750 mmHg.
At 300 ml/min dialysis
fl
uid
fl
ow maximum venous pressure during
treatment is approximately 250 mmHg.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:3
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...