The following will be shown on the Information Display when the
fi
rst
part of The Treatment Overview Menu is displayed:
QB
; current blood
fl
ow rate
P VEN
; current venous pressure
P ART
; current arterial pressure
UF RATE
; current ultra
fi
ltration rate
Activated functions are displayed at the bottom of the overview menu,
in the activity
fi
eld.
The second part of the overview shows the following on the
Information Display:
ACC QB
; accumulated blood volume since treatment start
ACC HEP
; accumulated infusion volume since treatment start
ACC UFV
; accumulated ultra
fi
ltration volume (patient weight loss)
since treatment start
UF RATE
; current ultra
fi
ltration rate
Activated functions are displayed at the bottom of the overview menu,
in the activity
fi
eld.
The
fi
rst part of the Treatment Overview Menu for single needle
treatment differs from the overview when performing double needle
treatment, showing parameters of interest when single needle mode is
active. Refer to chapter 5 in part 1.
3:24
AK 96
®
Operator's Manual - Operating the Machine
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...