alarm limits using the
Keypad
. See "Arterial Pressure Button" and
"Venous Pressure Button" in chapter 3 for instructions on manual
setting of the alarm limits.
The alarm windows will automatically be widened and the buttons
start to
fl
ash when the blood
fl
ow rate is changed or when the
Blood Pump button
is pressed.
Upon request from the clinic, an authorized technician can preset the
high arterial pressure alarm limit
to a
fi
xed value. This is to facilitate
pre-blood pump infusions during treatment during certain conditions,
see “Infusions during Treatment” page 3:36, part 1 for instructions.
Note that the function of the high arterial pressure alarm limit is in
this case deactivated implying that the operator is responsible for
monitoring the arterial pressure during treatment.
WARNING
When the high arterial pressure alarm limit is preset to a fixed
value
or when the arterial pressure measurement function is
not used, there is a risk for air-embolism if the arterial access is
accidentally disconnected. This is especially so when using the
machine in combination with a central venous catheter (due to
possible negative pressure in the catheter).
WARNING
TMP
When the
TRANSMEMBRANE PRESSURE CONFIRM MENU
is
displayed on the Information Display, press the
Select key
to con
fi
rm
the settings i.e. that the current TMP value and the automatically set
TMP alarm limits are set in accordance with the current UF rate and the
UF coef
fi
cient of the dialyzer being used. If the TMP alarm limits have
not been con
fi
rmed within two minutes, an attention appears where the
operator is requested to con
fi
rm the current settings.
If necessary the TMP alarm limits can be manually set/adjusted within
the preset alarm window by
fi
rst pressing the
Fluid Path button
,
selecting TMP, and adjusting the alarm limits using the
Keypad
. For
instructions on manual setting of the alarm limits, see "Fluid Path
button" in chapter 3. Go to "Transmembrane Pressure" last in the list.
The TMP alarm window will automatically be widened when the
UF Start/Stop button
is pressed.
CAUTION
When negative TMP alarm limits are set, the operator will not be
notified via alarm or attention alarm
that backfiltration may occur.
However, it is possible for the authorized technician to preset the
machine so that when negative TMP alarm limits have been set,
such an attention alarm occurs.
CAUTION
4:58
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...