INCORRECT DIA FLUID COMPOSITION
Check concentrates
Appears:
If the dialysis
fl
uid composition deviation (the relation
between the acidic and the bicarbonate concentrate) is
too high.
Machine actions:
The dialysis
fl
uid bypasses the dialyzer.
Measures:
Check concentrate selection and that the proper
concentrate has been connected. See "Connect/Con
Concentrates" on page 4:7 in part 1 for instructions.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:84
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order