After the venous pressure alarm has been attended to, and
if the venous pressure alarm window
is set to be wider
than the preset value, the
Venous Pressure button
will
start to
fl
ash. Press the
fl
ashing button to centralize the
alarm limits around the current value.
WARNING
When entering override, it is essential to remember
that the machine does not monitor that specific
parameter. The operator is responsible for the
monitoring during the override time.