BLOOD IN PRIMING DETECTOR
Function check stopped
Appears:
When the priming detector detects blood during function
check.
Machine actions:
The blood pump stops and the venous blood line clamp
closes.
Measures:
First, make sure that the patient has not been connected to
the blood lines. If this is not the case, clean the priming
detector lens. Alternatively call the authorized technician.
Mute time:
2 min.
Buzzer sound:
Continuous buzzer
14:16
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order