PROFILING IS NOT RESTARTED
Check settings and activate
Appears:
When treatment is resumed after the dialyzer and the
blood lines have been changed during treatment.
Machine actions:
The machine runs with the conductivity and/or the UF
rate settings present when
NEW BLOOD CIRCUIT
was
selected and discontinuing mode was activated.
Measures:
Check current set values for the previously activated
pro
fi
ling. reactivate Press the
Fluid Path button
, select
COND
and/or
UF
using the
Keypad
. Check that the start
and stop values for the pro
fi
ling type are the correct ones,
if not change the values. Reactivate the pro
fi
ling function
as usual procedure. See corresponding instructions in
chapter 7 "Pro
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:105