Measures:
-
Press the
fl
ashing
Alarm button
and the alarm text
will be shown on the Information Display.
-
Check cause for the low blood
fl
ow rate e.g. problems
with the patient's blood access that has recently been
taken care of, previously made air detector alarm
override, previous technical alarm etc. If possible,
increase the blood
fl
ow rate to the proper value and
the alarm will automatically disappear when the
blood
fl
ow rate is set above the alarm limit value.
-
If necessary, adjust the alarm limit value by pressing
the
Select key
to enter the
BLOOD FLOW MENU
.
Adjust the alarm limit;
SET LOW
, as necessary.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms
13:23