Kt/V alarm limit
The alarm limit for Kt/V is the lowest acceptable Kt/V
forecast value (calculated by the machine) accepted when
the treatment is complete. The value that is to be set is
the minimum Kt/V value which has to be achieved when
the treatment has
fi
nished. The alarm limit for Kt/V can
be set between OFF and 1.95.
To set the alarm limit for minimum Kt/V, select
Kt/V LIMIT
and set a proper alarm limit value, after
the distribution volume (
VOLUME
) has been set , using
the
Keypad
. If the Kt/V forecast value, estimated at that
time to be obtained when the treatment is complete, is
below the Kt/V alarm limit set, an attention alarm will
be generated:
Check cause, adjust treatment parameters or adjust the
alarm limit for Kt/V as above.