BiCart HOLDER NOT INSTALLED
Check concentrate selection
Appears:
BiCart concentrate has been selected, but the BiCart
holder has not been installed.
Machine actions:
None.
Measures:
Check selected concentrate. See "Connect/Con
Concentrates" on page 4:7 in part 1 for instructions.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:13
Attention alarms in alphabetical order