Step-by-step instructions for
measuring clearance and Kt/V
Single clearance measurement check
Procedure
1. Press the
Fluid Path button
,
select
DIASCAN
.
2. Select
SET DIASCAN
and the
setting menu will be displayed.
3. When
DIASCAN OFF
has been
selected, press the
Select key
.
4. Select
SINGLE
using the
Display Up key
. Single
measurement is now activated.
The single clearance measurement check will be
performed on an appropriate occasion chosen by the
machine. The occasion chosen is dependent on when
the self-calibrations of the machine are to be performed.
DIASCAN
will be shown at the activity
fi
eld of the
Treatment Overview Menu when the function has been
activated.
12:20
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx