Diascan MEASUREMENT FAILED
Press SELECT key to confirm
Appears:
The machine was not able to perform the measurement
check.
Machine actions:
If a single Diascan measurement check has been activated
it will automatically be deactivated; the Diascan function
is turned off.
If continuous Diascan measuring has been activated, the
current Diascan measurement check will automatically
be deactivated. The Diascan function will continue to be
on and the following measurement checks will continue
as set.
Measures:
Con
fi
rm the machine actions by pressing the
Select key
.
If desired, reactivate the single Diascan measurement
check.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Soft buzzer
14:40
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order