pH (option) Alarm
Appears:
The pH value of the dialysis
fl
uid is outside the set alarm
limits.
Machine actions:
-
The dialysis
fl
uid is bypassed from the dialyzer.
-
When the pH value of the dialysis
fl
uid is within the set
alarm limits again, the dialysis
fl
uid automatically enters
the dialyzer.
Possible cause:
•
The content of the concentrate container does not
correspond with the speci
fi
cations given.
•
Incorrect concentrate has been connected.
•
Either the acidic container or the bicarbonate
container/BiCart cartridge is empty (when using
bicarbonate concentrates).
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Alarms