Start the Treatment
CAUTION
Make sure that the blood path of the Flow Diagram lights up, if it
does not light up, check that the venous blood line is correctly placed
in the priming detector. If the blood path does not light up, the
machine considers the patient not connected and as a consequence
the UF system is not controlling the patient ultrafiltration.
If the air detector has not been activated during priming it will not be
automatically activated until the priming detector detects blood.
CAUTION
1. Check/adjust the blood
fl
ow rate (
SET QB
) to the
required value suitable for the particular patient using the
Blood Pump Up/Down keys
.
2. If desired, press the
Blood Path button
and open the
BLOOD FLOW MENU
using the
Keypad
. Check/adjust the
low alarm limit for blood
fl
ow rate in
SET LOW
as necessary.
3. Press the
fl
ashing
Arterial Pressure button
to centralize the
alarm limits around the current value. Or if desired, set the
arterial pressure alarm limits manually using the
Keypad
.
4. Press the
fl
ashing
Venous Pressure button
to centralize the
alarm limits around the current value. Or if desired, set the
venous pressure alarm limits manually using the
Keypad
.
5. Press the
fl
ashing
UF Start/Stop button
(lights up) to start
ultra
fi
ltration. Note especially that at all times whenever
necessary during treatment, the
UF Start/Stop button
can
be pressed (
fl
ashing) in order to decrease the UF rate to
minimum UF rate.
Check that the minimum UF rate is set to the correct value.
Press the
Fluid Path button
, select
UF
and check the value in
SET MIN UFR
. Adjust if necessary.
4:56
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...