Central Chemical Disinfection - Performing
The AK 96 dialysis machine can be programmed to perform central
chemical disinfection . This function, as well as the type of
disinfectant, has to be preset (not default) by an authorized technician.
In this mode, the machine receives disinfection solution through the
central water supply system. The central disinfection program in
AK 96 dialysis machine performs a continuous rinse through the
fl
uid
monitor until a preset time has passed or the operator has activated the
rinse/drain function.
It is advisable to turn the AK 96 dialysis machine off if it is not
supposed to be disinfected; this is to make sure that no disinfection
solution will enter the machine when not desired.
Note
•
The pick-up tubes must be rinsed/disinfected separately. See
"Pick-up Tubes” on page 8:41 for instructions.
Note
Check before starting the procedure
That the dialysis
fl
uid tubes are connected to the safety couplings
and that the latches of the BiCart holder are closed.
That the concentrate connectors are placed in the corresponding
stand-by ports of the machine (blue and red), and that the yellow
disinfectant connector is placed in the parking port (marked P).
Procedure
Use the
Keypad
next to the Information Display to select options and
to navigate within and between the menus. For details concerning the
keys included in the
Key pad
, see "Keypad" in chapter 3. part 1
1. Press the
Rinse/Disinfection button
brie
fl
y.
2. The
RINSE/DISINFECTION MENU
will be displayed on
the Information Display.
3. Select
CHEMICAL
.
4. Select the disinfectant that is set for CENTRAL
administration.
5. Start the program by pressing the
Select key
for 3 seconds,
Rinse/Disinfection button
lights up.
The machine now performs a continuous rinse through the
fl
ow
path of the machine.
8:32
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...