dialysis
fl
uid tubes must remain connected to the safety couplings until
the bypass path of the Flow Diagram lights up green.
When "F.Ch” disappears from the Time Display
the dialysis
fl
uid
tubes can be attached to the dialyzer.
The bypass path of the Flow Diagram also lights up. Before the correct
conductivity level has been reached the bypass path will be orange.
When the
fl
uid preparation is
fi
nished
the bypass path turns green
.
At the same time the
Fluid bypass button
will be
fl
ashing. When
pressed (and green path), the button lights up and the dialysis
fl
uid
will start to
fl
ow.
If the dialysis
fl
uid tubes are connected to the safety couplings, the
fl
uid will circulate within the machine. If they are connected to the
dialyzer, the
fl
uid will enter the dialyzer at that time. See "Dialysis
Fluid Tubes - Attach" starting on page 4:35 in part 1 for instructions on
how to properly attach the dialysis
fl
uid tubes to the dialyzer.
If desired, the concentrates may be connected, and the blood lines may
be attached, before the machine has been switched on. If so, do not
connect the pressure transducers of the blood lines to the machine until
the machine has been switched on and the
Priming button
lights up.
Priming may be started (the blood lines and the blood compartment of
the dialyzer) during function check (after complete blood part checks).
4:4
AK 96
®
Operator's Manual - Double Needle Treatment
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...