5. Check that the preset values in the model is correct. Activate
the pro
fi
ling model, by pressing the
Select key
for 3 seconds.
Or, if desired, it is possible to activate the pro
fi
ling function
at any time during priming or treatment. Bear in mind,
that the parameter settings of the complete model will be
activated simultaneously. If necessary, it is possible to
adjust the preset values of the parameters, refer to "Pro
without Preset Model", page 7:7.
Note
•
If UF pro
fi
ling has been activated
If UF volume or treatment time is changed, the pro
fi
ling graph
will automatically change as well. Always check the UF
pro
fi
ling settings if UF volume or treatment time is changed
after UF pro
fi
ling activation. For further details, see "Pro
ltration rate", page 7:3.
•
If sodium or bicarbonate pro
fi
ling has been activated
If treatment time is changed the pro
fi
ling graphs will not
change. As a consequence, the set stop values will not be
reached if treatment time is decreased.
•
If isolated UF is to be used
, and pro
fi
ling has been activated,
the pro
fi
ling will be bypassed during the isolated UF phase.
PROFILING
will still be displayed on the Information Display.
When the isolated UF phase is complete, the pro
fi
ling will
automatically be reactivated starting where it was interrupted.
•
If sodium or bicarbonate pro
fi
ling is deactivated during
treatment
, the machine will continue running on the values
from the point at which it was stopped. If pro
fi
ling is
reactivated without changed pro
fi
ling parameters, the machine
will continue running from the point where pro
fi
ling was
deactivated.
•
If UF pro
fi
ling is deactivated during treatment
, the machine
will recalculate the UF rate according to remaining UF volume
and treatment time. If UF pro
fi
ling is reactivated, without
changed pro
fi
ling parameters, the UF pro
fi
ling graph will
automatically change according to remaining UF volume and
treatment time. Always check the UF pro
fi
ling settings after
deactivating and reactivating UF pro
fi
ling.
•
If pro
fi
ling is activated (preset model) during treatment,
the
pro
fi
ling graph will still follow the preset parameters within
the remaining treatment time.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Profiling
7:11
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...