Alarms
Clearance alarm limit
The alarm limit for clearance is the lowest acceptable
clearance value measured during treatment.
To set the alarm limit for minimum clearance, select
K LIMIT
and set a proper alarm limit value
using the
Keypad
. The starting point value is 50% of the current
blood
fl
ow rate at that time. Adjust the value as necessary.
If the measured clearance value is below the alarm limit
set, an attention alarm will be generated:
Check cause, adjust treatment parameters to increase
clearance or adjust the alarm limit for clearance as above.
12:18
AK 96
®
Operator's Manual - Diascan
HCEN12239 Revision 12.2010
Program version 3.xx