FLUID PATH OBSTRUCTION
DURING DISINFECTION
Appears:
If a obstruction in the
fl
uid path has been detected during
disinfection.
Machine actions:
Disinfection is aborted and the dialysis
fl
uid pumps are
stopped. Drain is automatically performed when pressure
has decreased.
Measures:
Wait until drain is
fi
nished.
-
Check if the obstruction is external, i.e. the
fl
uid
tubes is clamped.
-
If equipped with UFD, check the UFD-
fi
lter.
Restart the disinfection. If the attention alarm recurs, call
for an authorized technician.
Mute time:
5 min.
Buzzer sound:
Soft buzzer.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:63