
ART PRESSURE LIMITS TOO WIDE
Press ARTERIAL button to adjust
Appears:
The arterial pressure alarm window
is wider than the
preset value.
Machine actions:
None
Measures:
Press the
fl
ashing
Arterial Pressure button
to centralize
the alarm limits around the current value.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
14:8
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...