Actions that will interrupt the measurement check are:
- the blood
fl
ow rate is changed
- the
Blood Pump button
is pressed
- the
UF Start/Stop button
is pressed
- the
Fluid Bypass button
is pressed
- a treatment alarm which produces dialysis
fl
uid bypass
or blood pump stop is generated.
- sodium and bicarbonate (conductivity) of the dialysis
fl
uid is changed
- temperature of the dialysis
fl
uid is changed
- dialysis
fl
uid
fl
ow rate is changed
- UF volume is changed
- linear UF pro
fi
ling settings is changed
- linear UF pro
fi
ling activation/deactivation is made
- sodium or bicarbonate pro
fi
ling activation/deactivation
is made
- isolated UF activation is made
In any of these cases the measurement check is
interrupted. The attention alarm is to be con
fi
rmed by
pressing the
Select key
and the machine will automatically
reschedule measuring as soon as possible.
Note
•
The blood flow rate and the dialysis fluid flow rate
have to be set within certain ranges in order for
correct Diascan measuring results to be obtained.
See chapter 9 "Technical Data and Specifications" in
part 1 for further details.
Note
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan
12:7
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...