standard procedure used by the clinic. The patient ends of the
arterial and venous blood lines must be connected together to
make a closed extracorporeal circuit. Use a disposable blood
line connection designed for this purpose, between the two
blood lines. In addition, connect priming
fl
uid via an infusion
line in a similar way as when infusions are administered to
the extracorporeal blood circuit during treatment. This is to
compensate for the priming
fl
uid being ultra
fi
ltrated from the
closed circuit. For information on how to connect the priming
fl
uid, see “Infusions during Treatment”, page 3:36 in part 1 of
this manual.
To enter recirculation priming procedure press the
Select key
when
RECIRCULATION
has been selected. The menu will change and
request the operator to press the
fl
ashing
Blood pump button
when
ready for recirculation priming. When the button is pressed, the
pump speed will automatically start at 150
ml/min. Adjust as
necessary using the
Blood Pump Up/Down keys
.
The recirculation priming overview menu, which gives current
information on the most important recirculation priming
parameters, will be shown on the Information Display. The
preset recirculation priming UF volume is shown below
PRESET UF VOLUME
. Below
ACC UF VOLUME
, to the left of
the slash, the current recirculation priming UF volume obtained
since recirculation start is shown. The total preset recirculation
priming UF volume is displayed to the right of the slash.
When the recirculation priming UF volume
and time
have
been achieved, the blood pump automatically stops.
EXTRA RECIRCULATION
can be restarted as desired. Check
the priming
fl
uid connected via the infusion line. Note that the
values below
ACC UF VOLUME
will be the accumulated ones
since the
fi
rst recirculating priming procedure.
Select
CONNECT PATIENT
using the
Keypad
when the
recirculation priming procedure is complete, and the menu will
change. Prepare the extracorporeal blood circuit for treatment in
accordance with standard procedure used by the clinic. Continue
by following the instructions in "Treatment", page 4:52 in part 1.
If the operator wishes to restart priming and at the same time set
the priming volume value to zero, select
NEW PRIMING
when
the priming volume achieved menu is displayed, using the
Keypad
. The machine will reactivate priming and it is possible to
repeat the priming procedure from the start.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Double Needle Treatment
4:51
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...