
Profiling with Preset Model
Selecting and Activating the Model
CAUTION
The operator has to make sure that the profile chosen is suitable for
the patient treated. The profiling parameters have to be checked
prior to the treatment.
Once the profiling values have been set, the operator should
verify that the control and protective systems for sodium and
bicarbonate profiling operate with the desired values and
that they are in agreement. This verification is made in two
steps. First by checking the set start (
START
) and stop
(
STOP
) values for
NA PROFILING
(protective values),
press the
Back key
and compare with the values
NA
displayed
in the
PROFILING MENU
(control values). Secondly by
checking the set start (
START
) and stop (
STOP
) values
for
HCO3 PROFILING
(protective values), press the
Back key
and compare with the values
HCO3
displayed in the
PROFILING MENU
(control values).
During UF profiling, the operator must readjust the TMP alarm limits
to follow the changing UF rate.
CAUTION
1. Press the
Fluid Path button
and select either
UF
or
COND
using the
Keypad
.
2. Select
PROFILING
.
3. Press the
Select key
and the
PROFILING MENU
will be
displayed on the Information Display.
4. Below
PRESET MODULES
, select
MODULE 1
,
MODULE 2
or
MODULE 3
using the
Keypad
.
7:10
AK 96
®
Operator's Manual - Profiling
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...