Conductivity
From the Fluid Path main menu; select
COND
by using the
Keypad
to
reach the sub menu.
The following will be shown on the Information Display:
SET NA
; set sodium value
SET HCO3
; set bicarbonate value
PROFILING
COND (C/P)
; current dialysis
fl
uid conductivity value, where C is
the current conductivity value from the control system, and P is the
current conductivity value from the protective system
SET COND (C/P)
; set dialysis
fl
uid conductivity value, where C is
the set conductivity value from the control system, and P is the set
conductivity value from the protective system
PH
; current dialysis
fl
uid pH value
Dialysis fluid flow rate
From the Fluid Path main menu; select
DIA FLUID FLOW
by using
the
Keypad
to reach the sub menu.
The dialysis
fl
uid
fl
ow rate is a preset parameter value which cannot be
adjusted by the operator during priming. The value can however be
set in (
SET TREATMENT DIA FLUID FLOW
) during priming, but
will not be effective until treatment start ( when the priming detector
detects blood in the venous blood line). In treatment the operator has
access to adjust the value.
The following will be shown on the Information Display:
SET TREATMENT DIA FLUID FLOW
; set dialysis
fl
uid
fl
ow rate
ACTUAL DIA FLUID FLOW
; current dialysis
fl
uid
fl
ow rate
Press SELECT 3 seconds to start CSBM
; this text is
used when activating concentrate stand-by mode. See "Concentrate
Stand-by Mode" page 3:30 in part 1 for instructions.
Diascan
®
From the Fluid Path main menu; select
DIASCAN
by using the
Keypad
to reach the sub menu. For instructions on settings and
activation of the Diascan function see chapter 12 in part 2.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Operating the Machine
3:17
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...