BLOOD PUMP STOP TIME EXPIRED
Start blood pump
Appears:
When the blood pump has been stopped for more than 1
min.
.
Machine actions:
None
Measures:
Start the blood pump.
Mute time:
2 min.
Buzzer sound:
Continuous buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:23
Attention alarms in alphabetical order