BPM button
Press the
BPM button
to reach the BPM main menu. On this menu
blood pressure measuring can be activated. For instructions see chapter
Fluid Bypass button
The
Fluid Bypass button
is used to manually bypass the dialysis
fl
uid
from the dialyzer.
UF Start/Stop button
The
UF Start/Stop button
is used to start and stop the ultra
fi
ltration.
Fluid Path button
Press the
Fluid Path button
to reach the Fluid Path main menu. In
this menu parameter settings and functions related to dialysis
fl
uid
can be set/activated.
The following will be shown on the Information Display:
UF
; calculated ultra
fi
ltration rate and pro
fi
ling type
CONC
; selected concentrate
TEMP
; current dialysis
fl
uid temperature
COND
; current conductivity value
DIA FLUID FLOW
; current dialysis
fl
uid
fl
ow rate
DIASCAN
; function activated or not activated
TMP
; current transmembrane pressure
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Operating the Machine