FCH IS WAITING
INLET WATER TEMP IS TOO HIGH
Appears:
If inlet water temperature is too high during function
check.
Machine actions:
Function check is interrupted. When inlet water
temperature has decreased the function check will
automatically continue.
Measures:
Check possible cause:
•
Inlet water temperature has not decreased below 45
°C within 5 minutes after the high temperature test
is started.
•
Inlet water temperature has not decreased below 39
°C within 5 minutes after the high temperature test
is
fi
nished.
If attention does not disappear, call for an authorized
technician.
Mute time:
Permanently muted buzzer.
Buzzer sound:
Super soft buzzer
14:52
AK 96
®
Operator's Manual - Attention Alarms
HCEN12239 Revision 12.2010
Program version 3.xx
Attention alarms in alphabetical order
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...