5. Start the program by pressing the
Select key
for 3 seconds,
the
Rinse/Disinfection button
lights up. Note that if the
automatic start of heat disinfection or rinse program has been
activated, the
Rinse/Disinfection button
will already be lit.
DISINFECTION IN PROGRESS:
will be displayed and will
remain throughout the program.
The
Rinse/Disinfection button
will continue to be lit throughout
the program, indicating that the program is ongoing.
6. The program starts with a rinse phase, therefore wait for
approximately 2-4 minutes before attaching the CleanCart
cartridge to the holder. An attention alarm will be generated
to notify you when, see next point.
7. When the attention alarm:
appears, open the latches of the BiCart holder and place the
CleanCart cartridge to the holder in a similar way as when
you place the BiCart cartridge to the holder.
Default, the cartridge can be left in the holder during the
entire program (up until next function check is started) and
will automatically be emptied. The program will now start,
continue to point 9 in these instructions.
By preset, an attention alarm can appear to request the
operator to empty the cartridge. In this case do point 8
before continuing to point 9.
8:14
AK 96
®
Operator's Manual - Hygiene and Maintenance
HCEN12239 Revision 12.2010
Program version 3.xx
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...