8. When
Kt/V TARGET
has been
selected set a desired dialysis
dose target value (Kt/V)
which
is to be achieved at treatment
end, using the
Display Up key
.
The target value for Kt/V can be
set between
OFF
and 2.0.
When the Kt/V target value has
been set, the machine calculates
the time needed to reach the
target. This time to target (
TTT
)
will be displayed in the viewing
menu as soon as the
fi
rst Diascan
measurement check has been
performed.
The
fi
rst Kt/V and clearance measurement check in
the sequence will be performed on an appropriate
occasion chosen by the machine. The occasion chosen is
dependent on when the self-calibrations of the machine
are to be performed.
DIASCAN
will be shown at the
activity
fi
eld of the Treatment Overview Menu when the
function has been activated.
During the period of time when the measurement check
is being performed,
DIASCAN
, shown at the activity
fi
eld
of the Treatment Overview Menu, will be
fl
ashing.
When a measurement check has been performed, an
attention alarm text without a buzzer will appear on the
Information Display . Press the
Select key
to enter
the
LOG
menu. The measurement result values are also
displayed in the Diascan viewing menu.
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Diascan
12:29
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...