Disinfection and Cleaning - Heat Disinfection
Temperature:
+93 °C (measured after heating rod)
≥
80 °C (measured in the outlet before the
heat exchanger)
Note
•
The temperatures are veri
fi
ed at nominal values for the mains
voltage and at 20 °C ambient temperature.
•
Total time for disinfection programs is estimated and may vary.
Note
One of three alternatives for heat disinfection can be selected. The
second alternative is presetable for a combined heating program and
the third is a CleanCart / heating alternative.
The default settings are as follows:
Heat disinfection
Total Time (min)
AK 96
Disinfection
programs
230 V
115 V
Heat
36
40
Heat Citric acid 20
%
52
57
With UFD
Heat CleanCart
50
55
Heat
33
37
Heat Citric acid 20
%
49
54
Without UFD
Heat CleanCart
46
51
Auto heat disinfection
Auto heat can be used with or without CleanCart. If the auto heat shall
be performed with CleanCart must the cartridge be installed before
the auto heat starts.
Total Time (min)
AK 96
Disinfection
programs
230 V
115 V
Heat
31
35
With UFD
Heat CleanCart
43
49
Heat
31
35
Without UFD
Heat CleanCart
43
48
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Technical Data
9:9
Summary of Contents for AK 96
Page 2: ......
Page 6: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 263: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 265: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 315: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...
Page 317: ...This page is intentionally left blank HCEN12239 Revision 12 2010 Program version 3 xx ...