BiCart HOLDER NOT CLOSED
Close or check BiCart holder
Appears:
When the BiCart holder is not properly closed.
Machine actions:
Buzzer is activated 2 min. after the attention alarm is
generated.
Measures:
Close the BiCart holder. See "Machine aftercare" on page
4:63 in part 1 for instructions.
Mute time:
2 min.
Buzzer sound:
Soft buzzer