Activated isolated UF is displayed at the activity
fi
eld of the
Treatment Overview Menu on the Information display. During
the phase, the accumulated values for ultra
fi
ltration
ACC UF
and
time
ACC TIME
will be displayed on the isolated UF menu.
Press the
Fluid Path button
, select
UF
and then
ISOL UF
to
view the menu.
If necessary, isolated UF can be deactivated at any time during the
phase, by pressing the
Select Key
for 3 seconds when the isolated UF
menu is displayed.
When the time for the isolated UF phase equals 0:00, the attention
alarm
TREATMENT TIME EXPIRED
appears and the treatment is to
be discontinued as usual procedure.
6:6
AK 96
®
Operator's Manual - Isolated UF
HCEN12239 Revision 12.2010