VENOUS PRESSURE LIMITS TOO WIDE
Press VENOUS button to adjust
Appears:
The venous pressure alarm window
is wider than the
preset value.
Machine actions:
None.
Measures:
Press the
fl
ashing
Venous Pressure button
to centralize
the alarm limits around the current value.
Mute time:
2 min.
Buzzer sound:
Soft buzzer
HCEN12239 Revision 12.2010
Program version 3.xx
AK 96
®
Operator's Manual - Attention Alarms
14:131
Attention alarms in alphabetical order